Project Description

Infusion pump CI-900

Infusion pump |#CI-900

Brief Introduction

FEATURES

●Patent technology of outside caliper rotary detector, avoid the pharmaceutical crystallizing and stove
pipe occasioned by traditional inner box slide drop sensor

●Three work modes: Drop counting control, volume control and time control mode

●Unique energy saving Li-ion battery inside, 6 hours continuous work in high capacity

●Unique colour LCD

●Special high for night operation

●Bolus function

Specifications

●Detector: Outside caliper rotary detector (Patient technology)

●Control way: Drop counting and adjusting the delivery rate automatically

●Pump type: Peristaltic finger

●Compatible IV sets: Compatible with IV sets of any standard (including lucifugal IV sets)

●Accuracy: ≤25drop/min ±2%, >25drop/min ±3%(drop counting and time control model);
±5%(volume control model with correct calibration)

●Delivery rate range: 1.0ml/h~1200ml/h (1.0ml/h~99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
1~400drop/min (1drop/min step)

●Delivery time range: 1~1999min (1min step)

●Volume limit range: 0~9999min (1ml step)

●Total volume range: 0~9999min (1ml step)

●Purge rate: 600.0ml/h

●KVO rate: KVO is 1ml/h when the delivery rate is 10ml/h, Otherwise KVO is 3ml/h

●Alarms: Air-in-line Infusion complete when volume limit has been set, Empty container, Occlusion, Door
is opened,  waiting over time, No drop been detected, Operational errors, Low battery, device failures

●Bolus rate: 10~1200ml/hr   Bolus volume: Maximum of 100ml

●Power source: AC power supply: AC 100~240V ±10%, 50/60 Hz  DC:12V, for Ambulance use

●Internal Battery: 11.1V, 2000mAh, 6 hours of continuous operation no a new fully charged battery with
●infusion running at 20 drop/min

●Power consumption ≤25VA

●Shock protection: Class I , Type CF, IPX4

Your Content Goes Here

Your Content Goes Here

Your Content Goes Here

  • Country Origin: China

  • Brand Name: VIOMED or OEM

  • Certification: CE, ISO, FDA

  • Payment: TT or L/C